Voltammetry slide, §5-1 Basic principle of polarograpy
本页为图片内容,未提取到文字,位于「Voltammetry」与「When applied v」之间,建议对照原课件图片补充描述。
When applied voltage slide, reaches the decomposition voltage of metal
查看本课件的文字版(共 110 页,供检索与复制)

第1页:Voltammetry

§5 Voltammetry§5-1 Basic principle of polarograpy Voltammetry group of analytical methods based on determining current flow–voltage curve during electrolysis Several types of methods based on Electrode type How the potential is applied How the current is measured.

第2页:

第3页:When applied voltage

When applied voltage reaches the decomposition voltage of metal ion:Reduction reaction on cathode:Cd2++ 2e-CdOxidation reaction on anode:2OH--2e H2O + 1/2 O2U外∝i Where U外denotes applied potential,Udis decomposition potential, is total resistance of electrolysis circuit,iis current through the circuit.U外-Ud= R(Cd2+)

第4页:When following conditions

When following conditions are obeyed, U外–i graph is showedin following Figure Lower current density during electrolysis Adequatelystirring to removeconcentration gradient No quantitative relationship between current (or potential)and concentration ofinterested ions

第5页:To find the analytical

To find the analytical relationship between the current andthe concentration of sought-for ions , following measurements are adopted:

MICROPLATINUM ELECTRODE Or DROPPING MERCURY ELECTRODE( in most common use)to insure high current density

NO STIRRIN Gto insure enough high concentration difference between the electrode surface and the solution bulk Then = Eo+ ln CMR Tn

第6页:When the current

When the current flowed for only short period through theelectrode, ion concentration on the surface of the electrodereduce suddenly. The difference in ion concentrations between the surface and the bulk solution is equivalent to an electro chemical cell, which is calledconcentrationpolarization(浓差极化).thus, there is voltage that is equivalent to this concentration change calledconcentrationpolarization potential(浓差电势)When microplatinum electrode or dropping mercuryelectrode is used and no stirring is carried out,concentration polarization takes place soon, followingpolarogram is obtained Limiting currentABCD

第7页:

第8页:When applied voltage does

When applied voltage does not reach reduction voltage of metal ion, only little current flow, called residual currentproduced by reduction of impurities and charging current,eg ABin the Figure Limiting currentABCD Arriving on decomposition voltage,the electrode reactions occur, the current rises rapidly with increasingvoltage,BCsection in the figure Cathode (DME):Cd2++ 2e-+Hg Cd(Hg)(amalgam)Anode (calomel or pool mercuryelectrode with big area):2Hg-2e-+ 2Cl-Hg2Cl2

第9页:When it reaches point

When it reaches point voltage, due to concentrationpolarization, the current arrives at limiting value anddoes not rise markedly, called limiting current,consisting of residual current and diffusion current极谱波可以用i~ U外表示曲线表示,也可以用i~ Ede曲线来表示,从下面的讨论可以看出,二者是基本重合的。U = ( ESCE-Ede) + R∵ and are very little inpolarographic electrolysis = ESCE-Ede=-Ede( vs. SCE)To remove completely influence of R, three electrodesystem is usually used.

第10页:Very little oscillation

Very little oscillation

第11页:

第12页:

第13页:Polarography First

Polarography First voltammetric technique Differs form hydrodynamic

unstirred (diffusion dominates)

dropping Hg electrode (DME) is used as aworking electrode Current varies as drop grows then falls off

第14页:

第15页:In polarographic analysis

In polarographic analysis, supporting electrolyte is usually added to increase the current-conductivity and eliminate migration current. Usually relatively higher concentration of strong electrolytes (alkali metal salts)serves as supporting electrolyte通常用通入N2,H2等方法消除溶解氧对极谱测定的影响。

第16页:In polarographic a(续)

(活塞)(螺线管)(金属垫圈))(聚氨酯)0.05~0.5mmdiameter(iii) Easy to removediffusion layer on mercury drop surface when the drop falls Hg microelectrodes

第17页:Advantages of DME

Advantages of DME (compared to planar electrode)

clean surfacegenerated

rapidachievement of constant current during dropgrowth

remixingof solution when drop falls toremovepreviousdiffusion layer

high Hgovervoltagemeans even metals with high–Ve E0can be measured without H2formation Disadvantages of DME

Hg easilyoxidized, limited use as anode(E < +0.4V)2Hg + 2Cl--2e→Hg2Cl2

non faradaic residual currentslimit detertion to > 10-5mol.L-1.

cumbersome(笨重的,尴尬的)to use(toxic mercury)

sometimes producecurrent maximafor unclearreasons( use maxima suppressor)

第18页:Sample Cells

Sample Cells

第19页:

第20页:

第21页:Equation of diffusion

§5-2 Equation of diffusion current basis ofpolarographically quantitative analysis On dropping mercury electrode:Mn++ ne-+Hg M(Hg)Ede= Eo+ lnRTnFCe CaWhere Ceis Mn+concentration on the electrode surface,Cais concentration in amalgam on electrode surface

第22页:Diffusion layer Bulk

Diffusion layer Bulk solution Electrode Diffusion layer Bulk solution Electroded Concentration polarizationaround mercury drop

第23页:When the applied voltage

When the applied voltage exceeds the decomposition voltage,diffusion-controlled current is expressed as:i= K(C-Ce)When the applied voltage gets more negative, Ce→0, thenid= KC Idreaches limiting value proportional to ion concentration in bulk solution, and do not changes withapplied voltage longer

第24页:In above equations

In above equations, is called Ilkovic constant, it is expressed as follows:Ilkovic equation---diffusion current equation = 607nD1/2m2/3t1/6id= 607nD1/2m2/3t1/6CAverage limiting diffusion current denoting average current on mercury drop from drop forming to falling (mA)Number of transferringelectrons in electrodereaction(e/mol)Diffusion coefficient of electro active analyte in solution(cm2.sec-1)Mercury mass flow rate(mg.sec-1)Drop time(sec)Concentration of electro-active analyte(mmol.L-1)From above equation, we can find thatwhentemperature, matrix solution and capillary characteristic are kept constant, idis proportional to

第25页:

第26页:

第27页:How it works

How it works?

The applied voltage is gradually increased, typicallyby going to more positive( more negativedecomposing potential)

small residual current is observed.

When the voltage becomes great enough, reductionoccurs at the analytical electrode causing current.

The electrode is rapidly saturated so current product ion is limited–based on diffusion of the analyte to the small electrode.

第28页:How it works

How it works ?The reduced species alters the surface of the mercury electrode.To prevent problems, the mercury surface isrenewed by “ knocking off ” drop–providing afresh surface.This results in an oscillation of the data as it iscollected.

第29页:Direct comparison method

Direct comparison method Polarographically quantitative analytical methods If t, m, and are constant, then we can simplify things by using standard.Then

第30页:Wave height

Wave height of polarographic wave is determined easierthan peak current, so following equation is in morecommon use

第31页:Calibration curve method

Calibration curve method

(or h)Chx CxPolarographic working curve As shown in right graph

第32页:Standard add it ion

Standard add it ion method When only one sample is analysized, this method can be employed.At first measuring the polarographic wave height hxofanunknown solutionwith volume of V, then alittleamount of the standard solution of concentration CSwitha volume of VSisaddedand the wave height of the mixture solution is measured. As we know, following equations are vivid:

第33页:在进行定量测定时

在进行定量测定时,通常只需测量所得极谱波的波高(以毫米或记录纸格数表示),而不必测量扩散电流的绝对值。

第34页:Quantitative Analysis

Quantitative Analysis

From Ilkovic equation:Id= KC Usually use method of standard Add it ions(with uLadditions, no volume correction needed)

Detection Limits:~ 10-5~ 10-6mol.L-1

Resolution:DE1/2≈0.2V(not very good)How do the DL and Resolution be improved?

第35页:Halfwave

§5-3 Halfwave potential—polarographic qualitative analysis下面推导极谱波方程式A + ne-BC Aeconcentration of on mercury drop surfaceCAconcentration of in bulk solution CBe concentration of the electrode surface if soluble, CBe concentration of in the solution near the electrodeif forms amalgam, then it means concentration in amalgam if is metal element and in soluble in mercury, then it equals to 1CBequals to zero, the concentration of in bulk solution

第36页:Nernst equation is still

Nernst equation is still vivid here:Ede= Eo+0.059gAC AengBC Belg Where-id= kACA-i= kA(CA-CAe)CAemay be calculated according to above eqs.CAe=-id+ ik

第37页:On the other hand

On the other hand, according Faradaic law, during electrolysis,concentration of reduction product ion is proportional to current flow, that is,CBe∝-iCBe=-KiSupposed = 1/ B-ikBHence Ede= Eo+0.059g AkBid-ing BkAilg•= Eo+0.059g AkBng BkAlg0.059id-in ilg+

第38页:E’ = Eo+

E’ = Eo+0.059g AkBng BkAlg令When = ½ id, log term in above equation is equal to zero,corresponding potential is called half wave potential E1/2E1/2= Eo+0.059g AkBng BkAlg

E1/2independent on the concentration

basis of qualitative analysis Ede= E‘+0.059id-in ilg(*)(#)

第39页:Because cathode current

Because cathode current is in more common use, equation (*) isrewriting as:For reduction wave Ede= E1/2+0.059 (id)c-icniclg Ede= E1/2-0.059 (id)a-ianialg For an oxidation wave Ede= E1/2+0.059 (id)c-ini–(id)alg For complex wave For three types of polarographic waves mentioned above, E1/2have same expression as formula(#)

第40页:Because cathode cu(续)

E1/2

第41页:

第42页:Because cathode cu(续)

i(mA)Ede(vs.SCE)VPolarographic waves of different concentrations of Cd2+[Cd2+] :in 1M KCl solution Why using E1/2but not Edecas qualitativeanalysis basis?

第43页:iEE1/2

iEE1/2

第44页:

第45页:

第46页:

第47页:

第48页:

第49页:Interference

§5-4 Interference currents and their removing methods

residual current

redox reactions of impurities in solution

charging of Hg drop(non-faradaic current / non-redox current)Charging current form at ion is shown in the figuresnext page

第50页:Interference(续)

≈+++++-----≈≈+++++-----GGG0.1MNull potential point++++++++DME Calomelelectrode Corresponding to section cb After voltage applied DME Calomelelectrode Corresponding to section ac Before voltage applied当接通回路,但未施加外加电压时,滴汞电极与参比电极短路,这时汞滴带正电,溶液中阴离子向电极表面扩散

第51页:

第52页:电容电流是由于汞滴表面与溶液间形成

电容电流是由于汞滴表面与溶液间形成的双电层,在与参比电报连接后,随着汞滴表面的周期性变化而发生的充电现象所引起的。此电流与滴汞电极的电位有关。对于微量组分(如<10-5M)的测定,虽然注意到所用试剂的纯度并经过仔细的除氧,但由于电容电流的存在,仍有微量的残余电流(约为10-7A数量级)通过(Why some 10-7A? Please refer topage 157~158, in the teaching book),这已足以引致困难。所以电容电流的存在是提高极谱分析灵敏度的主要障碍。残余电流一般可采用作固的方法予以扣除(见下图)。

第53页:Migration current

Migration current In polarographic analysis, you must removeconvection(对流)migration(迁移)of the sought-for ions in the solution Why?Migration current The current produced by static attractionof the electrode to sought-for ions它与被分析物质的浓度之间并无一定的比例关系,故应予以消除。

第54页:How to remove them

How to remove them?

Hold the solution still(使溶液保持禁止)to eliminate convection

Add supporting electrolyte to remove migration Supporting electrolyte:strong electrolyte, inert(electro-in active), with concentration 100 timesthan sought-for ion

第55页:Maximum

Maximum (or malformed peak )极大(或畸峰)在极谱分析中,常常出现一种特殊现象,即在电解开始后,电流随电位的增加而迅速增大到一个很大的数值,当电位变得更负时、这种现象就消失而趋于正常),这种现象称为极大或畸峰。++-----+++++++Shieldingregion(具体原因请参照教材P159)Measurements removing current maxima Add maximum suppresor, usuallysurfactant, eg.动物胶,聚乙稀醇,羧甲基纤维素, Triton X-100等,≤0.01%

第56页:Oxygen wave

Oxygen wave试液中的溶解氧在滴汞电极上被还原而产生两个极谱波:The first:O2+2H++ 2e-H2O2( acidic solution)O2+ 2H2O +2e-H2O2+ 2OH-(neutral or basic medium)E1/2=-0.05V(vs.SCE)The second:H2O2+ 2H++ 2e-2H2O (acidic medium)H2O2+ 2e-2OH-(neutral or basic medium)E1/2=-0.94V(vs.SCE)

第57页:Oxygen wave eliminating

Oxygen wave eliminating methods

Hydrogen wave

第58页:

第59页:Characteristics

§5-5 Characteristics and disadvantages ofpolarographic analysis

advantages•Relatively high sensitivity, kinear range:10-2~ 10-4mol.L-1•Relative error:±2%,as good as that of spectrophotometry•Simultaneously determine 4 ~ 5 analytes at one time underappropriate condition•A little amount of test sample needed•rapid analysis•Excellent reproductivity (通过的电流小,对溶液组成无明显影响)•Wide application field (metal ion, metal complex, an ion ororganic compounds etc thatis electro-active)

第60页:Disadvantages

Disadvantages•poor sensitivity•prewave interference• poor resolution,DE1/2≥100mVfor practical separation of two peaks

第61页:Polarographic Catalytic

§5-6 Polarographic Catalytic Wave Diffusion process Electrodereaction Chemical reaction Adsorption process Reactantproduct

第62页:如果化学反应与电极反应同时发生

如果化学反应与电极反应同时发生(平行),而且化学反应为定速步骤,则该极谱波称为极谱催化波,例如,A + ne-→ (electrode reaction)B + X→ + (chemical reaction)k1• is called catalyst• polarographic current is called catalytic current,i1∝[A]• chemical reaction is its velocity-determining step• must have high over-potential on the electrodeto insure catalytic circle

第63页:For example Total reaction

For example Total reaction: 2Fe2++ H2O2→ 2OH-+ 2Fe3+An other one:MoO52-2H++ 2e→ MoO42-+ H2OMoO42-+ H2O2→ MoO52-+ H2O

第64页:When no adsorption

When no adsorption exists, catalytic wave and classicalpolarographic wave have same shape, catalytic current is expressed as follows:i1= 0.51n D1/2m2/3t2/3k1/2Cx1/2CA Limiting catalytic current Concentration of and in solution Rate constant

第65页:For

For polarographiccatalyticwave, the current is independent on height of mercury column:i1∝m2/3t2/3∝h2/3h-2/3∝1Difference between catalytic wave and diffusion wave:But for diffusion-controlled currentid∝m2/3t1/6∝h2/3h-1/6∝h1/2Temperature coefficients:1% ~2% / ℃ forid4% ~ 5% and more / ℃i1催化氢波(自学)

第66页:Single-sweep polarography

§5-7 Single-sweep polarography(单扫描极谱法)(Oscillographic polarography,示波极谱法)Classical polarography:

Slow direct current voltage scanning rate: 0.2V/min, some100 drops of mercury per scan

Big residual current Single-wseep polarography:

Rapid scanning rate: 0.25V/sec, one drop of mercury per scan

Little residual current

Oscillographic polarograph

第67页:UDMER PtSC Eiip

UDMER PtSC Eiip Ep-Uiipand Epare respectively peak current and peak voltage is toothed wave voltage(锯齿波电压)

第68页:For reversible electrode

For reversible electrode reaction, diffusion current equation of single-sweep polarography is expressed as followsip= 2.69×105n3/2D1/2u1/2ACuis scan rate of voltage(V.s-1), is electrode area (cm2)Relationship between potential at peak Epand half wave potential E1/2Ep= E1/2–1.1 = E1/2-RTnF0.028n(25℃)For reduction wave, Epis-28/n more negative than correspondingE1/2, and for oxidative wave, 28/n more positive than E1/2

第69页:Advantages

Advantages of single-sweep polarography•Sensitivity: DL ~ 10-7mol.L-1, 2-3 magnitude order higherthan classical polarography•More precise, measuring peak height, but not wave height•Simple, rapid•Higher resolution, peak separation of 35 ~ 50m resolved•Little prewave interference, 5s interval time beforescanning•Little oxygen wave interference, in-reversible electrodereaction shows very weak or even no wave

第70页:BACK由于使用高输入阻抗的电位计

BACK由于使用高输入阻抗的电位计,所以极谱电流主要从对电极上通过,电解回路中流过的电流很小,因而工作电极的电位完全受外加电压所控制,参比电极的电位保持恒定

第71页:

第72页:

第73页:Back

Back

第74页:

第75页:Cyclic voltammetry

Cyclic voltammetry

施加等腰三角形脉冲电压

循环伏安法与示波极谱法具有同样形式的峰电流和峰电位方程式0.0-5.0+5.0ipaipcDEp EiCathodal wave Reduction(阴极)Anodic wave Oxidation(阳极)tUUi Us

第76页:The peak potential

The peak potential difference between cathodal peak and anodic peak is DEp= Epa–Epc= 2.22RTn F=56.5n(mV)对于不可逆体系,DEp > 56.5/n(mV),ipa/ipc< 1,阴阳峰电流比值越小,则该电极体系越不可逆

第77页:

第78页:

第79页:

第80页:

第81页:

第82页:Squar-wave polarography

§5-8 Squar-wave polarography

第83页:自学

自学

掌握方波极谱法提高极谱分析测定灵敏度的原理

方波极谱法有何特点?

应用方波极谱时应注意那些问题?

第84页:Pulse polarography

§5-9 Pulse polarography脉冲极谱法是在方波极谱法的基础上发展起来的,施加电压的方式也与方波极谱基本类似,即在线性扫描直流电压的基础上施加一个脉冲电压。不同的是方波极谱中方波电压是连续施加的,而脉冲电压是不连续的,即当汞滴增长到一定时间时,提供一个电压脉冲。这种方式可以显著地降低电容电流,从而提高测定的灵敏度、选择性。脉冲极谱法有两种类型:█差示脉冲极谱法:线性扫描电压+脉冲电压灵敏度最高,波形与方波极谱相似,测定峰值电流█常规脉冲极谱法:振幅随时间增加的脉冲电压波形与经典极谱波相似

第85页:

第86页:

第87页:

第88页:

第89页:自学

自学
掌握脉冲极谱法提高测定灵敏度的原理

第90页:Stripping voltammetry

§5-10 Stripping voltammetry溶出伏安法也称为反向溶出极谱法,这种方法是使被测定的物质,在适当的条件下电解一定的时间,然后改变电极的电位,使富集在该电极上的物质重新溶出,根据溶出过程中所得到的伏安曲线来进行定量分析。

第91页:

第92页:Stripping voltamme(续)

(沉积过程)

第93页:Stripping voltamme(续)

(溶出过程)

第94页:根据溶出过程中阳极电流与离子浓度的线性

根据溶出过程中阳极电流与离子浓度的线性关系进行定量

第95页:Electrodes in stripping

Electrodes in stripping voltammetry

Hung mercury electrode

glass carbon electrode mercury plating(镀汞膜的玻璃态石墨电极)Problems

Compared with other polarographic methods, how aboutthe sensitivity of sripping voltammetry?

what advantages do hung mercury electrode and mercury membrane electrode have ?

第96页:

第97页:

第98页:

第99页:

第100页:

第101页:单指示电极安培滴定

§5-11单指示电极安培滴定(Amperometric titrationwith single indicating electrode, polarographic titration)Amperometric titration:keeping potential constant Potentiometric titration:keeping current constant orat null-current伏安滴定法是应用伏安曲线的原理来确定化学计量点的容量分析方法。如果在滴定时观察电流的变化来确定化学计量点,就称为电流滴定或安培滴定.如果通过观察电位的变化来确定化学计量点,则成为电位滴定.

第102页:单指示电极安培滴定(续)

a: analyte is electroactive (able to be reduced on electrode),but the titrant is notb: analyte and titrant are all able to be reduced on the polarelectrode

第103页:Electrodes Reference

Electrodes Reference electrode:calomel electrode,mercury pool electrode Working electrode:dropping mercury electrode,solid microelectrode Advantages• Solid microelectrode can be used• wide linear range. 0.1 ~ 10-4mol.L-1Using electrode with big area, DL can be down to10-6mol.L-1Back titration can increase upper limit of determination• some electro-in active materials can also be measured using electro-active titrant.

第104页:Example Th4+Al3+Fe3+

Example Th4+Al3+Fe3+(added)Titrated by F-ThF62-AlF6-Fe3+(added)Titration continued ThF62-AlF6-FeF63-(非电活性)

第105页:双指示电极安培滴定

§5-12双指示电极安培滴定(永停滴定)(Dead-stop end point titration)下图

第106页:由于外加电压很小

由于外加电压很小,只有可逆电对才有电流通过

不可逆电对,电解池中无电流通过,如左图中1和5Ce4++ Fe2+= Ce3++ Fe3+

当01时,溶液中分别存在Fe3++ e→Fe2+和Ce4++ →Ce3+可逆电对,要使电解回路中产生电流,需要的外加电压很小,当a=0.5,a=1,5等时,电对的可逆性最高,而当a=0或a=1时,电对为完全不可逆。

在永停滴定时,外加电压很小,所以只有可逆电对才有电流产生。

第107页:iaABCO Titration curve

iaABCO Titration curve of dead-stop end point method

With reversible system titratingreversible one, curve ABCE in the leftfigure. Eg. Ce(IV)→Fe3+

With irreversible system titratingreversible one, curve ABCD in the leftfigure, eg. S2O32-→I2

With reversible system titratingirreversible, curve ABC in the lower left side. Eg. I2→ S2O32-(永停法)

第108页:双指示电极电位滴定

§5-13双指示电极电位滴定(potentiometric titration with double indicating electrodes)

potentiometric titration includes two kinds single-indicating electrodedouble-indicating electrode

双指示电极电位滴定(恒电流)和双指示电极电流滴定(恒电位)相似。区别在于,后者保持电位恒定,观察电流的变化,前者则保持电流恒定,而观察电极电位的变化。

第109页:

第110页:对于可逆电对

对于可逆电对,产生同样大小的电流所需要提供的电压小于不可逆电对IIIIIII Va01滴定终点可用作图法求出,也可直接由pH计指针的摆动确定。a =0a =1a=0.5a>1←DE→←DE→